Image Dimension Measurement System

IM-6000 series

This model has been discontinued.
Compliance with the certification standard is ensured as of the time of shipment from our company.

Measurement Head: Wide-field model IM-6120

IM-6120 - Measurement Head: Wide-field model

*Please note that accessories depicted in the image are for illustrative purposes only and may not be included with the product.

Data Sheet (PDF)

Manuals

  • CE Marking

Specifications

Model

IM-6120

Type

Measurement head

Image sensor

1" 6.6 mega pixel CMOS

Display

10.4" LCD monitor (XGA: 1024 × 768)
External monitor connectable (clone output)

Receiver lens

Double telecentric lens

Image measurement

Field of view

Wide field: ø100 × L200 mm, High precision: 25 × L125 mm

Minimum display unit

0.1 µm

Repeatability

W/o stage movement

Wide field: ±1µm High precision: ±0.5µm

With stage movement

Wide field: ±2µm High precision: ±1.5µm

Measurement accuracy (±2σ)

W/o binding

Wide field: ±5µm*1 High precision: ±2µm*2

With binding

Wide field: ± (7 +0.02L)µm*3 High precision: ± (4 +0.02L)µm*4

External remote input

No-voltage input (with contacts, without contacts)

External output

OK/NG/FAIL/MEAS.

Relay outputRated load 24 VDC 0.5AON resistance 50 mΩ or less

Interface

LAN

RJ-45 (10BASE-T / 100BASE-TX / 1000BASE-T)

USB 2.0 series A

6 ports (front: 2, rear: 4)

Record

Hard disk drive

250 GB

Illumination system

Transparent

Telecentric transparent illumination (Green LED)

Ring

Four division ring illumination (White LED)

External

Externally connected illumination (coaxial)

X stage

Moving range

100 mm (electric)

Withstand load

2 kg

Z stage

Moving range

30 mm (electric)

Power supply

Power voltage

100 to 240 VAC 50 / 60 Hz

Power consumption

310 mA or less

Environmental resistance

Ambient temperature

+10 to +35 °C

Relative humidity

20 to 80 % RH (No condensation)

Weight

Approx 31 kg

*1 In the range of ø80 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position
*2 In the range of ø20 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position
*3 In the range of ø80 × 180 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units)
*4 In the range of ø20 × 120 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units)

Data Sheet (PDF) Other Models